Advanced Search
    TIAN Shuxing, YAN Jiangsu, ZHOU Chongfu, LIU Hongguang, LI Junshan. Period verification of electronic balance in drug testing laboratories and uncertainty assessment of its indication error[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART A:PHYSICAL TESTING, 2023, 59(12): 74-76. DOI: 10.11973/lhjy-wl202312018
    Citation: TIAN Shuxing, YAN Jiangsu, ZHOU Chongfu, LIU Hongguang, LI Junshan. Period verification of electronic balance in drug testing laboratories and uncertainty assessment of its indication error[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART A:PHYSICAL TESTING, 2023, 59(12): 74-76. DOI: 10.11973/lhjy-wl202312018

    Period verification of electronic balance in drug testing laboratories and uncertainty assessment of its indication error

    • In order to ensure the normal operation of the electronic balance in the drug testing laboratory and the accuracy and reliability of the test data, the electronic balance in the drug testing laboratory was checked during the period according to the JJG 1036-2008, and the uncertainty of the indication error was evaluated according to the JJF 1847-2020. The results show that the verification results were qualified, the expanded uncertainty was 2. 5×10-4, the electronic balance could detect the sample normally.
    • loading

    Catalog

      Turn off MathJax
      Article Contents

      /

      DownLoad:  Full-Size Img  PowerPoint
      Return
      Return