Advanced Search
    ZHANG Qing-yun, ZHANG Jie, LU Ye-hang, LI Zhong-cheng. Application of Full Factorial Experiment Design in Residual Stress Measurement of High Stress Standard Sample by X-ray Diffraction Method[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART A:PHYSICAL TESTING, 2015, 51(7): 470-473. DOI: 10.11973/lhjy-wl201507005
    Citation: ZHANG Qing-yun, ZHANG Jie, LU Ye-hang, LI Zhong-cheng. Application of Full Factorial Experiment Design in Residual Stress Measurement of High Stress Standard Sample by X-ray Diffraction Method[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART A:PHYSICAL TESTING, 2015, 51(7): 470-473. DOI: 10.11973/lhjy-wl201507005

    Application of Full Factorial Experiment Design in Residual Stress Measurement of High Stress Standard Sample by X-ray Diffraction Method

    • A DMAIC flow of 6σ was made to study the influence of the parameters such as aperture diameter,exposure time and number of β angle on residual stress measurement of high stress standard sample by X-ray diffraction method.The results of full factorial experiment showed that the parameters had an great effect on amount of X photon,intensity of X-ray diffraction peak and ration of diffraction peak to background.On the premise of meeting the conditions of detecting efficiency,small aperture diameter,long exposure time and large number of β angle could increase accuracy and decrease repeatability error of the test results.According to the high stress standard sample,choose 2 mm aperture diameter,3 s exposure time and 11β angle could meet the requirements of the related standard value.