    电子拉力试验机横梁高速移动故障的维修

    REPARATION OF ABNORMAL RUNNING FRAME OF ELECTRON TENSION TESTING MACHINE IN HIGH SPEED

    • 摘要: 针对INSTRON公司的电子拉力静态试验机出现横梁无法高速移动的故障,结合该试验机的基本结构和工作原理,逐一对可能引起故障的因素进行了检查和分析。结果表明,故障是由驱动试验机横梁的功率放大器损坏所造成的。维修后,试验机恢复正常工作。

       

      Abstract: Due to abnormal running of electron tension static testing machine frame of INSTRON corporation in the high speed,and based on the structure and system work theory of the testing machine,each factors possible caused breakdown had been checked and debugged. As result,a damaged power amplifier of drive machine caused frame breaking down. After maintence,the test machine recover work again.

       

    /